The handpieces and attachments must be lubricated
after each use to:
•
Increase their lifetime by reducing the wear between
moving parts.
•
Reduce heat generation by reducing the friction
between moving parts.
Failing to lubricate the parts will lead to damage and mal-
function, increasing the risk of harm to the user and pa-
tient.
For further information, please refer to the UNIUM Care
and Maintenance Poster and the Instruction for Use of
Synthes Special Oil (519.970, 519.97*).
Accessible moving parts of the handpiece and attach-
ments must be lubricated after each use with one (1) drop
of Synthes Special Oil (Fig. 1). Once the oil has been ap-
plied, move (e.g. press, open/close) the moving parts sev-
eral times to properly spread the oil. Wipe off excess oil
with a cloth. This applies to the following components:
Lubricating the Handpieces (Figs. 2 and 3)
•
Trigger shafts (05.001.601 and 05.001.611).
•
Attachment release buttons (only 05.001.601).
•
Mode switch (05.001.601 and 05.001.611).
•
Attachment coupling (05.001.601), reciprocating saw
blade coupling and release knob (05.001.611).
Lubricating the Casing Cover and Front Lid
•
Casing opening button (Fig. 4).
•
Shaft on the inside of the casing cover lid (Fig. 5).
•
The entire length of the sealing evenly (Fig. 6).
Wipe off excess oil with a cloth.