■ Notes:
•
The manual cleaning steps should also be followed to
clean the inside of the UNIUM Modular or
Reciprocating Saw Handpiece and the UNIUM
Washing and Sterilization Baskets, Insert Tray and Lids.
•
It is recommended to hold the UNIUM Modular
Handpiece and Reciprocating Saw Handpiece in an
upright position (Fig. 1) whenever possible and to drain
the inside of the Handpiece from excess water after
each step.
•
Handle the handpieces’ front lid, casing cover, edges,
seals and hinges with care.
1. Remove Debris
Rinse the device under cold, running tap water for a
minimum of 2 minutes (Fig. 1). Use a soft lint-free cloth or
soft- bristled brush to assist in removing gross soil
(Fig. 2 and Fig. 3).
For cannulations of the handpiece and attachments, the
cleaning brushes (519.40, 519.40* and 532.024) below
should be used.
2. Manipulate Moving Parts
Manipulate all moving parts such as triggers, sliding
sleeves, attachment release rings, saw blade coupling and
switches, etc. under running tap water to loosen and
remove gross debris.
Manual Cleaning Instructions