Disassembly
Before cleaning, remove all attachments and cutting tools
from the Power Tool. Remove the Power Unit from the
UNIUM Modular Handpiece and Reciprocating Saw
Handpiece.
Cleaning and Disinfection of the Power Unit, Adapter
for UBC II and UBC II Directly After the Surgery:
1. To clean the Power Unit, Adapter for UBC II and the
UBC II, wipe them off with a clean, soft and lint-free
cloth dampened with deionized water until no visible
soil remains on the devices. If any visible soil remains,
repeat the step with a new damped cloth. Dry them off
using a clean, soft, lint-free single-use cloth prior to
disinfection (Fig. 1 and 2).
2. To disinfect the Power Unit, Adapter for UBC II and the
UBC II, wipe them off with a new, clean, soft and lint-
free cloth dampened with a minimum of 70% alcohol-
based disinfectant for thirty (30) seconds. A
disinfectant that is VAH (Verbund für Angewandte
Hygiene) listed, EPA (Environmental Protection Agency)
registered or locally recognized is recommended. This
step has to be repeated two (2) additional times using a
new, clean, soft and lint-free cloth dampened with a
minimum 70% alcohol-based disinfectant each time.
▲ Precautions:
•
Do not use solvents to disinfect the Power Unit.
Sockets must not contact water or solvents as there is
a danger of short-circuiting.
•
Do not spray the contacts. When using a damp cloth to
wipe off the contacts, ensure that they are completely
dry before inserting the Power Unit into the handpiece.
Residual moisture could cause a short-circuit.
•
Inspect the Power Unit and Adapter for UBC II for
cracks and damage.
Place the Power Unit by using the Adapter for UBC II in
the UBC II after each use (Fig. 3).
Once the Power Unit is completely charged (indicated
by the green LED illuminating), remove it from the
Adapter and wipe it with a minimum 70% alcohol-based
disinfectant prior to returning to use.
Reprocessing of the UNIUM Handpieces, Sterile
Cover, Attachments and Accessories Directly
After the Surgery
•
All devices must be reprocessed using:
•
manual cleaning, or
•
automated cleaning with manual pre-cleaning
■ Note:
Clean all movable parts in opened or unlocked position.
Care and Maintenance
Cleaning and Disinfection
Figure 1
Figure 2
Figure 3
Preparation Prior to Reprocessing
29
UNIUM
™
• Instruction for Use