SV 971A User Manual
74
11 SOUND EXPOSURE METER
– DOSIMETER
The instrument operates as a Sound Exposure Meter (SEM) or Dosimeter in a very similar way to the
Sound Level Meter (SLM) and, in addition to the SLM results, measures also basic dose parameters.
This chapter describes settings specific for the SEM.
11.1 S
ELECTING
D
OSIMETER FUNCTION
To select the
Dosimeter
function, enter the
Function
section, select the
Meas.
Funct
position and press the
<Enter>
key. In the
Meas. Funct
screen, select the
Dosimeter
function and press the
<Enter>
key.
<ENT>
11.2
S
ETTING GENERAL PARAMETERS
–
G
ENERAL
S
ETTINGS
Most general settings of the
Dosimeter
function are similar to the
Level Meter
function (see Chapter
). Additionally,
Dosimeter
has a programmable five
automatic pauses.
Programable automatic pauses
Automatic pause(s) can be switched off (
Off
) or can be programmed based on
the RTC time (
On
).
If
Pause
is
On
, two additional positions appear which
enable setting time for pause begin (
Pause
Begin
) and
time for pause end (
Pause End
).
▼
Note: In the Simple interface mode, the Pause parameters are hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Off)
.
11.3 S
ETTING PROFILE PARAMETERS
–
P
ROFILE
X
Parameters of three profiles can be set in the
Profile x
lists of parameters.
The following parameters can be programmed
independently for each profile: weighting filter
(
Filter
),
peak filter (
Filter Peak
) and LEQ detector type
(
Detector
), criterion level (
Crit. Level
), threshold level
(Thr. Level
), exchange rate (
Exch. Rate)
, thresholds -
ULT Thresh.
and
PTC Thresh.
.
<ENT>
Weighting filter selection
Z
according to IEC 61672-1:2013 for Class 1,
A
according to IEC 651 and IEC 61672-1:2013 for Class 1,
C
according to IEC 651 and IEC 61672-1:2013 for Class 1,
B
according to IEC 651 for Class 1.
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...