SV 971A User Manual
– Appendixes
209
Table C.1.2.
Linear operating ranges for the
Low
measurement range (for the sinusoidal signal and
microphone sensitivity 32 mV/Pa)
[dB]
L
AS/F
L
BS/F
L
CS/F
L
ZS/F
L
AeqT
L
BeqT
L
CeqT
L
AE
L
Cpeak
(t
int
= 2 s)
from
to
from
to
from
to
from
to
from
to
from
to
from
to
from
to
from
to
31.5 Hz
24
83
24
106
24
120
32
123
24
83
24
106
24
120
27
86
50
123
500
Hz
24
119
24
122
24
123
32
123
24
119
24
122
24
123
27
122
50
126
1 kHz
24
123
24
123
24
123
32
123
24
123
24
123
24
123
27
126
50
126
4 kHz
24
124
24
122
24
122
32
123
24
124
24
122
24
122
27
127
50
126
8 kHz
24
122
24
120
24
120
32
123
24
122
24
120
24
120
27
125
50
123
12.5 kHz
24
118
24
117
24
117
32
123
24
118
24
117
24
117
27
121
50
120
Note: For the signals with the crest factor n > 1.41 upper measuring range of the RMS (LEQ
and SPL) is reduced. The valid upper limit can be calculated according to the below given
formula:
, where A is the upper limit for the sinusoidal signal
Example: For the crest factor n = 10 the upper limit is A10 = 120 dB
Measuring ranges
Reference measuring range
of the acoustic pressure
Normal
Measuring frequency range
of the acoustic pressure
(-3 dB) 5
Hz ÷ 20 000 Hz.
Basic measurement error
of the acoustic pressure
< 0.7 dB
(measured for the reference
conditions, see below).
Weighting filters
(see Chapter C.3)
Z
meeting requirements of the IEC 61672-1:2013 standard for the Class 1
“
Z
“ filter
A
meeting requirements of the IEC 651 and IEC 61672-1:2013 standard for the Class
1 “
A
” filter
B
meeting requirements of the IEC 651 and IEC 61672-1:2013 standard for the Class
1 “
B
” filter
C
meeting requirements of the IEC 651 and IEC 61672-1:2013 standard for the Class
1 “
C
” filter
Table C.1.3.
Self-generated noise for different weighting filters
Electrical *)
Acoustical compensated
Weighting filter
Range
A
C
Z
A
C
Z
Low
< 12 dB
< 12 dB
< 20 dB
< 17 dB
< 17 dB
< 23 dB
Normal
< 16 dB
< 16 dB
< 23 dB
< 20 dB
< 20 dB
< 27 dB
*) measured with the
ST 03
microphone equivalent impedance
18 pF@100kOhm
± 10%
Special filters
•
Diffuse Field
compensation filter that improves the complete instrument
frequency response in the diffuse acoustic field (see below)
•
Windscreen
compensation filter that improves the instrument frequency
response in the free acoustic field when windscreen SA 22 is
mounted on the microphone (see Chapter C.1.2)
( )
2
log
20
137
n
A
n
−
=
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...