![Svantek SV 971A User Manual Download Page 37](http://html1.mh-extra.com/html/svantek/sv-971a/sv-971a_user-manual_1409270037.webp)
SV 971A User Manual
37
Gradient trigger
The
G
trigger starts the 1-second measurement/integration under the
condition: value of the RMS result (
Source
) integrated during 0,5 ms is greater
than the threshold (
Level
) and the gradient of the Source value is greater than
the gradient threshold (
Gradient
). In other cases, the instrument continues
checking the trigger condition every 0.5 mc.
This type of trigger has the same logic as the
Level+
trigger, but the trigger condition requires also
gradient level to be exceeded.
Note: When a measurement is waiting for the gradient trigger, the
flashing “trigger” icon superimposes on the „wait” icon.
Source result
Only one measured result (
Source
) can be used for checking trigger condition in
the
Level Meter
mode, namely the instantaneous LEQ from the first profile (with
appropriate filter and detector), which is denoted here as
Leq(1)
. This position
cannot be changed.
Threshold level
The threshold (
Level
) can be set in the range from
24 dB
to
136 dB
. The
Source
value compares with the
Level
value every 0.5 milliseconds.
Speed of Source value changing
This position appears when the
G
trigger is chosen. The speed of the
Source
value changing (
Gradient
) can be set in the range from
1 dB/ms
to
100 dB/ms
.
4.3 S
ETTING PARAMETERS FOR PROFILES
–
P
ROFILES
Parameters for three profiles can be set in the
Profiles
screen (in case of
Lev. Met.
function) or in the
Profile x
screens (in case of
Dosimeter
function).
Following
parameters
can
be
programmed
independently for each profile: weighting filters
for other
than peak results
(
Filter
), weighting filters for peak
results (
Filter Peak
) and LEQ detectors type
(
Detector
).
<ENT>
Weighting filter
Next weighting filters for both
Filter
and
Filter Peak
positions can be selected:
Z
according to IEC 61672-1:2013 for Class 1,
A
according to IEC 651 and IEC 61672-1:2013 for Class 1,
C
according to IEC 651 and IEC 61672-1:2013 for Class 1,
B
according to IEC 651 for Class 1,
LF
low frequency filter according to China requirements.
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...