SV 971A User Manual
46
The
Field Compensation
position
enables you to
switch off (Off) the compensation filter or to select the
compensation filter for sound measurements in the free-
field (
Free Field
), diffuse field (
Dif. Field
) conditions or
for the permanent outdoor monitoring application as a
part of the
SV 271
monitoring station:
Environment
(when the acoustic signal is parallel to the microphone’s
grid) or
Airport
(when the acoustic signal is
perpendicular to the microphone’s grid).. By default, it is
set to
Free Field
.
◄
The
Windscreen
position is active when the field
compensation is on and switches on the compensation
when the windscreen is applied. By default, it is
On
◄
Note: In the Simple interface mode, the Comp. Filter position is hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Microphone:
On; Field Comp.: Free Field; Windscreen: On)
.
Note:
For laboratory instrument’s tests, the compensation filters should be set in accordance
with the Appendix C
.
4.8 S
ETTING STATISTICAL LEVELS
–
S
TATISTICAL
L
EVELS
In the
Stat. Lev.
screen, you can define ten statistical
levels, named from
N1
to
N10
, to be calculated,
displayed and saved in a file as Summary Results (see
Appendix D).
Default statistical levels have following settings:
1
,
10
,
20
,
30
,
40
,
50
,
60
,
70
,
80
and
90
. All values should be
within the integer range [1, 99]. Each value can be set
independently from others.
<ENT>
4.9 P
ROGRAMMING INSTRUMENT
’
S INTERNAL TIMER
–
T
IMER
The
Timer
function is used to programme the automatic start-up of the measurement at a given time and
day of a week and with the parameters set in the
Measurement
section.
Note: In the Simple interface mode, the Timer position is hidden, but the instrument will use
settings previously defined in the Advanced mode or default settings (Timer: Off)
.
The
Timer
position allows you to programme the internal
real-time clock to act as a delayed start timer. The
instrument will be switched on automatically at the
programmed time and will perform the measurement
with the same settings used before the instrument was
turned off with one exception (see below Note).
<ENT>
Note: When Timer is On, measurements will be performed from defined Start to Stop times
because the Repetition Cycles parameter will be changed to Inf (path: <Menu> /
Measurement / General Set.). The last integration may be cut.
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...