elapsed. The instrument checks the trigger condition also during the recording
and if the condition is met the recording will continue for another
Rec. Time
.
Integration period trigger
When the
Integr. Per
trigger is selected, a signal recording is triggered every time
the measurement starts, and recording will last minimum
Rec. Time
. If the trigger
condition appears during recording (when
Integration Period
is shorter than
Rec. Time
), from this moment, the recording will continue for the next
Rec. Time
and so on.
Manual trigger
When the
Trig.manual
trigger is selected, the signal
recording starts and ends after pressing simultaneously
the
◄
and
►
keys during the measurement. After
pressing these keys, the screen with the corresponding
message appears. The registration always stops after
period defined by the
Rec
. Time
parameter.
Note: When a signal recording is waiting for the gradient trigger manual
trigger or “integration period” trigger, the flashing “trigger” icon
superimposes on the grey „signal” icon.
The
Filter
position enables the user to choose the broadband frequency filter
during a wave recording:
Z
,
A
,
C
,
B
or
LF
.
The
Sampling
parameter defines the sampling frequency of wave recording:
24 kHz
or
12 kHz
.
The
Signal Gain
position enables the user to choose the gain of the recorded
signal:
0 dB
…
40 dB
.
.
Source result
Only one measured result can be used as a trigger source
(
Source
) for checking
the trigger condition in the
Level Meter
mode, namely the instantaneous LEQ
from the first profile averaged for the
Trigger Period
with appropriate filter and
detector, which is denoted here as
Leq(1)
. This position cannot be changed.
Threshold level
The threshold (
Level
) can be set in the range from 24 dB to 136 dB.
Checking the trigger condition
The
Source
value compares with the
Level
value with the
step defined by the
Tr. Period
parameter which may be
set as:
Logger Step
,
0.5ms
,
100ms
and
1s
.
►
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...