SV 971A User Manual
79
12.3
S
TARTING
RT60
MEASUREMENTS
Measurements using Decay method
1.
Set parameters for
Decay
RT60 measurements. Most used setup is as
presented below.
•
Method:
Decay
•
Recording Time:
Auto
•
Time Step:
2ms
•
Averaging:
On
•
Noise Mar.:
10.0dB
•
Level:
100dB
.
2.
Place the sound power source in the measured room (for the sound power
source location - see the reverberation time measurement ISO standard).
3.
Place the microphone in one of the selected measurement points (for the
measurement points location see the reverberation time ISO standard).
4.
Switch on the sound power source.
5.
Start the measurement process by pressing the
<Start>
key. While the instrument is waiting for the
trigger condition fulfilment the
LCpeak
result is
displayed.
6.
Switch off the sound power source (the source
should work enough long to obtain the acoustic
field stabilisation). After the trigger condition
fulfilment, the instrument collects data and gives
the RT60 table for octave or third octave bands.
∕
Note: It is necessary to switch on the sound source before starting the measurement because
of the trigger requirements (for more details see Appendix E). If it is necessary to start the
instrument before switching on the sound source it is recommended to use the higher
Start Delay value.
Measurements using Impulse method
1.
Set parameters for
Impulse
RT60 measurements. Most used setup is
presented below.
•
Method:
Impulse
•
Recording Time:
Auto
•
Time Step:
2ms
•
Averaging:
On
•
Noise Mar.:
10.0dB
•
Level:
100dB
2.
Place the microphone in one of the selected measurement points (for the
measurement points location see the reverberation time measurement ISO
standard).
Note: The proper value of the sound level trigger threshold should be set well above
the background noise and significantly below the maximum sound level emitted by the
impulse source.
3.
Start the measurement process with the
<Start>
key. While the instrument is waiting for the trigger
condition fulfilment the
LCpeak
result is displayed.
4.
Release the impulse sound power source. If the
trigger condition is fulfilled the instrument collects
data and gives the RT60 table for octave or third
octave bands.
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...