SV 971A User Manual
88
2. To start measurement, press the
<Start>
key. The
Measurement
view
appears,
showing
the
measurement progress bar and STI index calculated
with considered ambient noise (
STI
), measured STI
index (
Meas.
), STI qualification band next to the
index,
LAeq
and
LAS
results.
The measurement lasts 15 seconds during which
the displayed results are updated every 1 second.
<Start>
Note: Ambient noise distortions will be considered in accordance with the values set in the
Ambient noise table
– see Chapter
Note: Pressing the <Stop> key during the measurement will break it without recording the
measurement results.
After completing the measurement, the instrument
shows the
Point
view with current measurement
results. You can scroll measurement results with the
▲
/
▼
key.
3. Press the
<Start>
key if you wish to perform another
measurement.
or
4. To finish measurement series for the measurement
point, you should complete them.
To complete the measurement series in the
measurement point, press the
<Enter>
key and in
the command list, select
Complete
and press
<Enter>
.
<ENT>
The instrument closes the series of measurements
in the current measurement point and switches to the
Area
view.
<ENT>
Note: In case of automatic averaging the series of measurements will be completed
automatically after meeting the conditions specified in the selected standard.
If you select the measurement in the
Point
view and open
the command list, you can view the modulation function
coefficients (
Modulation
) for the selected measurement
and delete (
Delete
) or exclude from the averaging
(
Exclude
) the selected measurement.
<ENT>
If you choose
Modulation
and press
<Enter>
the
instrument will go to the
Modulation function
view.
To change columns in this view, use the
◄ / ► key.
To return to the
Point
view press
<Enter>
or
<ESC>
.
<ENT>
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...