SV 971A User Manual
– Appendixes
296
The graphical illustration of data registering in this method is presented in Figure 7.
Fig 7. Data recording in the impulse method of the reverberation time evaluation
The measurement time in this method consists of:
•
The time before marker t
trig
in which the SVAN xxx analyser measures the actual sound pressure level
and waits for the very high impulse sound pressure level which will fulfil the trigger condition. The trigger
conditions will be fulfilled only when emitted impulse has maximal sound pressure level higher than L
t
level (cf. Fig. 6). The L
t
level in the SVAN xxx analyser is adjusted by parameter
Level
(
path: <Menu> /
Measurement / RT60 Settings
).
•
The time between markers t
s
and t
trig
registered since the trigger condition fulfilment back till point t
s
to
allow recognising the beginning of the decay region. In the SVAN xxx instruments this time is equal to
the
Time Step
(
path: <Menu> / Measurement / RT60 Settings
) parameter value multiplied by 50.
•
The time between markers t
trig
and t
e
registered since t
trig
forward to record whole decay curve together
with significantly long period of the noise level. This time in SVAN xxx instruments is adjusted by
Recording Time
(
path: <Menu> / Measurement / RT60 Settings
) parameter.
The above graph shows that the proper setting of the
Recording Time
value is very important. The
registration time has to be long enough to acquire sufficient number of background noise level values. In
other case the decay curve region could not be properly analysed or decay region could not fulfil the dynamic
condition mentioned above. It is recommended to set the
Recording Time
parameter two times longer than
expected reverberation time.
t [s]
L
[dB]
t
e
L
2
t
trig
10 dB
L
1
registration time
t
s
L
t
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...