Reference conditions
as per IEC 61260-1:2014
•
Reference temperature
+23
°C
•
Reference relative humidity
50%
•
Static pressure
101.325 kPa
Calibration
(electrical)
Calibration level
0.5 V
RMS
Basic accuracy
<
0.2
dB (for the temperature T=+23°C
5°C for sinusoidal
signal 120 dB
RMS
in the band 5
Hz ÷ 20 kHz with filter
Z
)
Measurement error
in the full temperature range
<
0.1 dB (when the temperature is from -
10°C to +50°C for the
sinusoidal signal 120 dB
RMS
in the band 5
Hz ÷ 20 kHz with
filter
Z
)
Overload detector
The instrument has the built-in overload detectors. Both A/D converter and input amplifier overload conditions
are detected. The overload in the measurement channel (in its analogue part) and the overload of the
analogue /
digital converter are both detected. The “overload” indication appears when the input signal
amplitude is 0.5
dB above the declared “Peak measurement range”.
Warm-up time
1 min. (for 0.1 dB accuracy).
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C re Reference conditions).
Effect of magnetic field
< 15 dB (A) or < 25 dB (Z) (for 80 A/m and 50 Hz).
Effect of Vibration
< 0.1 dB (from 20 Hz to 1000 Hz at 1 m/s
2
).
Antialiasing filter
Built-in antialiasing filter. Second-order analogue filter, passive Class, combined with on-chip FIR digital filter
of the analog-to-digital converter, ensuring correct sampling of the measured signal.
Pass band (-1 dB)
22.200 kHz
Pass band (-3 dB)
23.520 kHz
Stop band
26.256 kHz
Attenuation in the stop band
> 80 dB
Sampling frequency
48 kHz
(internal only)
Analogue to digital converter
1 x 24 bit
Reference range
Normal
Input attenuator accuracy
0.1 dB (for f = 1
kHz and T = +23°C)
Internal oscillator accuracy
0.01 % (for f = 1
kHz and T = +23°C).
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...