SV 971A User Manual
47
Measurement start and stop
The
Start (hh:mm)
and
Stop (hh:mm)
positions
determines times of measurement’s `start and stop. `
►
In the positions:
Monday
,
Tuesday
, …,
Sunday
; you can select days in a week
when measurements should start.
The timer can be programmed for
Max days
ahead (
up
to 100) or without
limitation (
Inf
) and during these days, the instrument refers to the time of the
R
eal
T
ime
C
lock (
RTC
). The measurement series stops when the day-counter number
will be equal to
Max days
. If
Inf
value is selected the measurement series can
be stopped only manually (if the power is assured). If more than one day in a
week is selected, every performed measurement will increase the day-counter.
After programming the Timer, you can turn off the instrument and it will turn itself
on before each measurement. However, there is a minimum time between turning
off the instrument and the set time in the Timer, which is 10 minutes. If you try to
turn off the instrument during this time, a question will appear whether you really
want to turn off the instrument, because it will also turn off the set Timer.
Note: Make sure to check that the real-time clock settings are correct before using the timer.
Note: Make sure that there is sufficient internal batteries power available for the instrument to
carry out the required measurements when it wakes up.
4.9.1 Example of timer execution
Let us assume that you wish to switch on the measurement on Monday at 14:54,
to measure noise level for 10 minutes and save results in the files with names
R1, R2, R3 etc.
To do this configure the
Timer
function as on the attached screen and to set the
measurement parameters (
path: <Menu> / Measurement / General Settings
) and
the file name (
path: <Menu> / Measurement / Logging / Logger Setup
).
The instrument will start to warm up during 30 seconds before the measurement
start time 14:54 on the nearest Monday.
The measurement will be performed by a period of ten minutes. Then, the results
will be saved in the file with the name R1 automatically and the instrument will be
waiting for the next Monday to start measurement at 14:54. Next file will be
automatically named R2 and so on. Such measurement will be repeated so many
times as was defined by the
Max days
parameter.
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...