SV 971A User Manual
26
File manager
File Manager
is used for checking content of the
memory and operations on files and directories such as:
renaming, deleting, displaying information and creating
new directories.
<ENT>
The
SD Card
memory is organised as a standard
memory with directories and sub-directories (FAT32 file
system). It is possible to create or delete directories.
The content of the memory can be checked with the help
of the
File Manager
position in the
File
section.
◄
Automatic saving of logger and wave files
Logger files are created and saved automatically to the SD-card. To enable automatic saving several
conditions should be fulfilled:
1. SD-card should be inserted and there should be enough free space on it.
2. The
Logger
(
path: <Menu> / Measurement / Logging / Logger Set.
) and/or
Recording
(
path:
<Menu> / Measurement / Logging / Wave Rec.
) should be enabled.
3. The new file should be defined with a unique name (
path: <Menu> / Measurement / Logging /
Logger Set. / Logger Name
and
path: <Menu> / Measurement / Logging / Wave Rec. / File Name
).
Files are saved in the directory, which was set as a working directory. The default working directory (after
using
Factory Settings
function) is called
SVANTEK
.
Note:
During the measurement run with data logging to the logger file, the “logger” icon is
displayed. When the signal is recorded to the wave file, the “signal” icon is displayed.
The file name (Logger or Wave) is generated automatically using a pattern
LLdd
,
where
LL
is the string of letters (so called
prefix
) and
dd
is a string of digits that
forms a number. Up to 8 characters can be used to name a file.
The default prefix for the logger files is
L
and for the wave files -
R
.
The instrument assigns an individual counter to each prefix of files the user has
created and saved in the working directory. The counter value is equal to the
maximum number in the set of files with the same prefix. For example, if there
are files with names:
L0
,
L15
and
L16
, the counter value is 16.
The number of the new automatically created file will have the value of the
counter increased by one. So, for the above example, new file name will be
L17
.
You can change the automatically generated file name
in the special screen, which is opened after pressing the
◄ / ► key.
After changing the number in file name without changing
the prefix and pressing
<Enter>
, the counter will be
automatically adjusted.
►
The instrument accepts only that name which number is
higher than the counter of the prefix.
<ENT>
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...