13.3.3 Results averaging
STIPA measurements require averaging of STI/CIS indexes. This averaging concern two aspects:
1) averaging of results for a measurement point.
Averaged
STI
/
CIS
for the measurement point is
displayed in the first line of the
Point
view and in the
measurement points list of the
Area
view.
2) averaging of results for a tested area.
Averaged
STI
/
CIS
for the tested area is displayed in the
second line of the
Area summary
view.
According to most standards, it is good practice to average the results for measuring points from two or
three measurements. In case of fluctuating noise occurring during measurements, it is recommended to
make not less than three measurements and check whether the spread of the obtained STI results does
not exceed 0.03.
The standard deviation is determined with the calculation of the mean STI (or CIS) for the area.
13.3.4 STIPA settings
STIPA measurements are configured in the
STIPA Settings
and
Ambient noise
screens. To configure
a STIPA project follow these steps:
1. Select
the
STIPA Sett.
position
in
the
Measurement
section and open it with the
<Enter>
key.
<ENT>
2. Being at the
Project
position, press the ► key and,
in the
File Manager
screen, select
<New Dir.>
to
create a new project directory or select the directory
of the previously created Project and press
<Enter>
.
►
In case of
<New Dir.>
the instrument will propose a
default name for the new Project directory which can
be modified in the editor screen opened after
pressing the
<Enter>
key.
After edition of the directory name it should be
confirmed by the
<Enter>
key.
<ENT>
Being at the new created directory line in the
File Manger
screen press
<Enter>
to return to the
STIPA Sett.
screen.
The selected project name will be displayed in the
Project
position of the
STIPA Sett.
screen.
<ENT>
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...