SV 971A User Manual
– Appendixes
211
RMS detector
•
Digital
“True RMS“ with Peak detection,
•
Resolution
0.1 dB
•
Range
327.7 dB
•
Crest Factor
unlimited (for signals in 20 kHz band).
Overload detector
The instrument has the built-in overload detectors. Both A/D converter and input amplifier overload conditions
are detected. The overload in the measurement channel (in its analogue part) and the overload of the
analogue / digital converter are both detect
ed. The “overload” indication appears when the input signal
amplitude is 0.5
dB above the declared “Peak measurement range”.
Underrange detector
The instrument has the built-in under-
range detector. The “underrange” indication appears when the Leq value
for the elapsed time or the last second L
XY
value is below the lower linear operating range.
Time
weighting
characteristics
(Exponential averaging)
Slow
“
S
“ according to IEC 61672 Class 1, Equivalent Time Constant 1000 ms
Fast
“
F
” according to IEC 61672 Class 1, Equivalent Time Constant 125 ms
Impulse
“
I
” according to IEC 60804 Class 1, Equivalent Time Constant 35 ms, Hold Time 1500 s
Reference conditions
as per IEC 61672-1:2013
•
Class of the acoustic field
Free field
•
Reference acoustic pressure
114.0 dB (related to 20
Pa)
•
Reference frequency
1000 Hz
•
Reference temperature
+23
°C
•
Reference relative humidity
50 %
•
Reference static pressure
1013 hPa
•
Reference incidence direction
perpendicular to the microphone diaphragm.
Calibration
Acoustical
-
with
the SV 36 sound calibrator
(or equivalent):
•
Calibration level for the pressure field
114.0 dB (equal to the calibrator pressure Level
– see
calibration chart of the used calibrator)
•
Calibration level for the free field and
0 deg incidence angle
114.0 dB (equal to the calibration level for the pressure field
minus free field correction of ACO 7152 at 1000 Hz
Note:
The above levels correspond to 114 dB of calibrator’s sound pressure. If the calibrator
has a different sound pressure than 114 dB, the calibration levels must be accordingly adjusted.
Maximum peak voltage
20 V Peak-Peak (Maximum peak voltage of input sinusoidal
signal, which can be applied to the SLM without destruction the
meter)
Warm-up time
1 min. (for 0.1 dB accuracy)
Summary of Contents for SV 971A
Page 220: ...SV 971A User Manual Appendixes 220 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 221: ...SV 971A User Manual Appendixes 221 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 222: ...SV 971A User Manual Appendixes 222 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 223: ...SV 971A User Manual Appendixes 223 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 224: ...SV 971A User Manual Appendixes 224 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 232: ...SV 971A User Manual Appendixes 232 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Page 233: ...SV 971A User Manual Appendixes 233 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 234: ...SV 971A User Manual Appendixes 234 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Page 235: ...SV 971A User Manual Appendixes 235 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 236: ...SV 971A User Manual Appendixes 236 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Page 298: ...SV 971A User Manual Appendixes 298 O ring gasket locations inside...