E
En
ng
glliis
sh
h
E
EN
N
1-54
8001-009-005 REV D
www.stryker.com
Alarm conditions
Remove from use mode
The Remove From Use (RFU) is a safety mode to limit operations. A fault condition prevents the product from normal
functions and requires service. The controller will stop the Active Therapy and communicate to the operator that the
controller is going into RFU mode.
CAUTION
Always remove the product from use before servicing any components. Contact qualified service personnel for service.
Depending on the remove from use (RFU) condition, text may or may not be displayed. For example, if there is a power
loss.
•
Water temperature probes are out of the allowed range
•
Program and data checksum error
•
High thermal cutout test failed
•
Backup power product replacement required
•
Low or over safety temperature
•
Pump over current
•
Compressor power fault
•
Heater power fault
•
Refrigerant control valve fault
•
Main DC power lost
•
CAN heart beat lost
•
Dual safety temperature sensors do not match readings
•
Dual safety temperature sensors are out of the allowed range
•
Hardware watchdog heartbeat failure
Summary of Contents for Altrix 8001
Page 2: ...sample text ...
Page 6: ...sample text ...
Page 68: ...sample text ...
Page 70: ...sample text ...
Page 138: ...sample text ...
Page 206: ...sample text ...
Page 208: ...sample text ...
Page 276: ...sample text ...
Page 344: ...sample text ...
Page 412: ...sample text ...
Page 414: ...sample text ...
Page 480: ...sample text ...
Page 482: ...sample text ...
Page 546: ...sample text ...
Page 610: ...sample text ...
Page 612: ...sample text ...
Page 682: ...sample text ...
Page 744: ...sample text ...
Page 746: ...sample text ...
Page 820: ...sample text ...
Page 883: ...0086 2016 08 D 2 8001 009 005 REV D www stryker com Altrix 精密温度管理系统 8001 操作 维护手册 ...
Page 884: ...sample text ...
Page 904: ...简 简体 体中 中文 文 Z ZH H C CN N 14 18 8001 009 005 REV D www stryker com 简介 生产日期 生产日期是序列号的前四位数字 ...
Page 943: ...0086 2016 08 D 2 8001 009 005 REV D www stryker com Altrix 精密温度管理システム 8001 取扱 保守説明書 ...
Page 944: ...sample text ...
Page 1005: ...0086 2016 08 D 2 8001 009 005 REV D www stryker com Altrix 정밀 체온 관리 시스템 8001 작동설명서 ...
Page 1006: ...sample text ...
Page 1067: ......