Alarm conditions
Patient probe malfunction alarm
This alarm notifies the operator that the probe is not providing information to the product during an active therapy.
Alarm generation:
When the primary patient probe is in a shorted, opened condition, or out of range for more than 30 seconds, the product
will display the Patient Probe Malfunction Alarm.
Note:
The product will deactivate the heat exchange and keep the pump activated as requested by the Active Therapy.
Patient probe disconnect alarm
This alarm notifies the operator that the probe is not providing information to the product during an active therapy.
Alarm generation:
When the adapter cable for the primary probe is removed and the reading of the Primary Patient probe is out of range for
more than 30 seconds, the product displays the Patient Probe Disconnected alarm.
Patient temperature deviation medium alarm
This alarm notifies the operator that the patient is not responding as expected in the active therapy.
Alarm generation:
The product will display the Patient Temperature Deviation Medium Alarm, if after the initial achievement of the current
patient temperature target, the actual primary patient temperature becomes 0.5° C or more above, or below the Current
Target Temperature.
Patient temperature output deviation alarm
This alarm notifies the operator that the patient temperature output is out of range or their is a calibration error.
Alarm generation:
The product will display the Patient Temperature Output Deviation Alarm, if the calibration has failed, or the patient
temperature output is out of range.
Normothermia deviation alarm
This alarm notifies the operator of that the Primary Patient Temperature is out of range.
Alarm generation:
If the actual Primary Patient Temperat ure is lower or equal to 35.9° C or higher or equal to 38.1° C, the controller will
display the Patient Normothermia Deviation alarm.
Water temperature deviation alarm group
This alarm notifies the operator that the water is not responding as expected to the therapy. The product is at full power,
with the current mode and the temperature selection. The water temperature cannot remain within a range of ±0.8º C of
the selected Water Target Temperat ure.
Summary of Contents for Altrix 8001
Page 2: ...sample text ...
Page 6: ...sample text ...
Page 68: ...sample text ...
Page 70: ...sample text ...
Page 138: ...sample text ...
Page 206: ...sample text ...
Page 208: ...sample text ...
Page 276: ...sample text ...
Page 344: ...sample text ...
Page 412: ...sample text ...
Page 414: ...sample text ...
Page 480: ...sample text ...
Page 482: ...sample text ...
Page 546: ...sample text ...
Page 610: ...sample text ...
Page 612: ...sample text ...
Page 682: ...sample text ...
Page 744: ...sample text ...
Page 746: ...sample text ...
Page 820: ...sample text ...
Page 883: ...0086 2016 08 D 2 8001 009 005 REV D www stryker com Altrix 精密温度管理系统 8001 操作 维护手册 ...
Page 884: ...sample text ...
Page 904: ...简 简体 体中 中文 文 Z ZH H C CN N 14 18 8001 009 005 REV D www stryker com 简介 生产日期 生产日期是序列号的前四位数字 ...
Page 943: ...0086 2016 08 D 2 8001 009 005 REV D www stryker com Altrix 精密温度管理システム 8001 取扱 保守説明書 ...
Page 944: ...sample text ...
Page 1005: ...0086 2016 08 D 2 8001 009 005 REV D www stryker com Altrix 정밀 체온 관리 시스템 8001 작동설명서 ...
Page 1006: ...sample text ...
Page 1067: ......