14. RESECTION MEASUREMENT
5. If measuring two or more known points, repeat
procedures 4 in the same way from the second
point.
6. Press
[CALC]
or
[YES]
to automatically start
calculations after observations of all known points
are completed.
• Instrument station elevation and standard
deviation, which describes the measurement
accuracy, are displayed in <Result>.
Standard deviation values for each point are
displayed in <Detail>.
7. If there are problems with the results of a point,
align the cursor with that point and press
[OMIT]
.
“OMIT” is displayed to the right of the point.
Repeat for all results that include problems.
8. Press
[RE CALC]
to perform calculation again
without the point designated in step 7 The result is
displayed.
If there are no problems with the result, go to step
9.
If problems with the result occur again, perform the
resection measurement from step 4.