21. CHECKS AND ADJUSTMENTS
PROCEDURE Checks and adjustments
1. Carefully level the instrument.
2. Position the prism in a horizontal direction
approximately 50 meters from the instrument.
3. Select "Inst.cons." in <Settings>.
4. Select "Image Sensor-Reticle offset".
5. Use manual sighting to accurately sight the target.
"10.3 Manually Sighting the Target"
• Sight not the center of the target plate but the
center of the prism.
6. Press
[OK]
.
• Press
[STOP]
to stop the measurement.
Prism center