Setup
3.4 Inputs and outputs
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
63
●
Medium temperature
●
Kinematic viscosity
All process values are available for analog output channel 2 and frequency output channel 4.
The accuracy specified for the analog output signal applies only within the range 4 to 20 mA.
Lower limit (4 mA) and upper limit (20 mA) can be assigned to any specific flow values.
①
Linear control range
②
Lower saturation limit
③
Upper saturation limit
④
Lower fault current value
⑤
Recommended setting range for lower fault current
⑥
Recommended setting range for upper fault current
⑦
Measuring range
Figure 3-1
Current limits for NAMUR configuration
The fail safe current output signal can be selected to:
●
Lower fault current (defined in the Loop current scale selection)
●
Upper fault current (defined in the Loop current scale selection)
●
Last valid value (the last process value before the failure occurred)
●
Current value (actual measured value)
●
Fail-safe value (within the range of 0 mA to 25 mA
1)
)
1)
For channel 1 the range is 3.5 mA to 25 mA
Output scaling configuration
Below are four examples describing configuration possibilities for a current output.
Summary of Contents for FST020
Page 2: ......
Page 122: ...Setup 3 14 Status icons FST020 Function manual 120 Function Manual 09 2017 A5E42949526 AA ...
Page 186: ...Security 6 9 Logout FST020 Function manual 184 Function Manual 09 2017 A5E42949526 AA ...
Page 188: ...Language 7 1 Language FST020 Function manual 186 Function Manual 09 2017 A5E42949526 AA ...
Page 192: ...Index FST020 Function manual 190 Function Manual 09 2017 A5E42949526 AA ...
Page 193: ......