Setup
3.2 Process values
FST020 Function manual
48
Function Manual, 09/2017, A5E42949526-AA
3.2.4.2
Custom units
Description
User specific text string to define a name for custom unit values. The custom unit can then
be selected in the sound velocity units.
Setting
Default
Related
Path
Setup
→
Process values
→
Sound velocity
→
Custom units
3.2.4.3
Custom conversion factor
Description
Conversion factor for custom unit sound velocity values based on SI velocity units m/s.
Setting
Default 1.0
Related
Path
Setup
→
Process values
→
Sound velocity
→
Custom conversion factor
3.2.4.4
Decimal places
Description
Sets the decimal places for the process value displayed in the local display.
Setting Auto; No decimal place; 1 decimal place; 2 decimal places; 3 decimal places; 4 decimal
places; 5 decimal places; 6 decimal places
Default 2 decimal places
Related
Path
Setup
→
Process values
→
Sound velocity
→
Decimal places
3.2.4.5
Alarm and warning limits
3.2.4.5.1 Upper alarm limit
Description
If process value exceeds this limit an alarm is triggered.