Setup
3.12 View 5
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
113
Related
This parameter is only available if View is set to Six Values or Six Diagnostic Values.
Path
Setup
→
Display
→
View 5
→
6th Process Value
3.12.9
Trend scale mode
Description
The Trend Scale Mode can be selected with automatic scaling or fixed scaling.
Setting Auto; Fixed
Default Auto
Related
This parameter is only available if View is set to Totalizer or 1 Value and Graph.
If Fixed is selected, Trend Scale Lower Limit and Trend Scale Upper Limit must be
configured.
Path
Setup
→
Display
→
View 5
→
Trend Scale Mode
3.12.10
Trend log time window
Description
The Trend log time window logging period (time axis length) can be selected.
Setting 1 Minute; 5 Minutes; 15 Minutes; 30 Minutes; 1 Hour; 2 Hours; 3 Hours
Default 5 Minutes
Related
This parameter is only available if View is set to Totalizer or 1 Value and Graph.
Path
Setup
→
Display
→
View 5
→
Trend log time window
3.12.11
Trend scale lower limit
Description
The Trend Scale Lower Limit defines the scaling lower limit of the value axis for the fixed
mode.