Setup
3.1 Sensor settings
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
31
Related
Path
Setup
→
Sensor settings
→
Path settings
→
Path 1
→
Serial number
3.1.7.3
Path diagnostic alarm limits
3.1.7.3.1 Percent bursts accepted alarm limit
Description
Sets the desired limit for the percentage of accepted bursts. If measured percentage
accepted burst drops below this limit for any installed path, an alarm will be issued.
Setting Min: 0%; Max: 100%
Default 50%
Related
Path
Setup
→
Sensor settings
→
Path settings
→
Path diagnostic alarm limits
→
Percent bursts
accepted alarm limit
3.1.7.3.2 Correlation factor alarm limit
Description
Sets the desired limit for the correlation factor. If the measured correlation factor drops below
this limit for any installed path, an alarm will be issued.
Setting Min: 0.0; Max: 1.0
Default 0.95
Related
Path
Setup
→
Sensor settings
→
Path settings
→
Path diagnostic alarm limits
→
Correlation factor
alarm limit
3.1.7.3.3 RxGain alarm limit
Description
Sets the limit for the RxGain value. If the RxGain goes above this limit for any installed path,
an alarm will be issued.
Setting Min: 0.0 dB; Max: 100.0 dB
Default 50.0 dB
Summary of Contents for FST020
Page 2: ......
Page 122: ...Setup 3 14 Status icons FST020 Function manual 120 Function Manual 09 2017 A5E42949526 AA ...
Page 186: ...Security 6 9 Logout FST020 Function manual 184 Function Manual 09 2017 A5E42949526 AA ...
Page 188: ...Language 7 1 Language FST020 Function manual 186 Function Manual 09 2017 A5E42949526 AA ...
Page 192: ...Index FST020 Function manual 190 Function Manual 09 2017 A5E42949526 AA ...
Page 193: ......