Instructions for Use Carina SW 3.n
49
Preparation
Internal battery
If the external power supply (mains/external bat-
tery) should fail, Carina automatically switches over
to supply from the internal battery.
If the battery is fully charged, ventilation can be per-
formed reliably for approx. 60 minutes without any
external power supply (assuming average ventila-
tion parameters and normal ambient conditions).
A completely discharged battery can be fully
charged within a charging time of approx. 2 hours.
Refer to "Maintenance of the internal battery" on
page 110.
The LED
lights up.
WARNING
Before transporting the device within the
clinic check that the internal battery is suffi-
ciently charged. Otherwise the correct func-
tioning of the Carina may be compromised.
CAUTION
Recharge the battery if it has not been used for a
lengthy period! Refer to "Maintenance of the inter-
nal battery" on page 110.
NOTE
If charging and discharging cycles are incomplete,
the charge capacity of the internal battery maybe
displayed inaccurately.
For improving the displayed accuracy, see
page 82.
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...