Preparation
52
Instructions for Use Carina SW 3.n
Connecting to the nurse call system
Socket on the back of Carina to transmit alarms to
a central alarm system within the clinic.
Plug: AMP/Tyco 5-555237-3, 6-pin
Pin allocation:
View of the wiring side of the open plug
1
N.O. contact (closes in the event of an alarm)
2
Base
3
N.C. contact (closed during normal operation)
Dräger makes the following recommendations for
safety reasons: Use an N.C. contact so that any
disconnection or cable break will trigger a nurse
call.
z
Plug into the » « socket on the rear of the
device.
z
Check that the connected nurse call system is
operating correctly.
Technical data of the nurse call system
WARNING
A nurse call system used in conjunction with
Carina must conform to IEC 60601-1-8 and
transmit alarms from Carina within 5 seconds.
Risk of injury to patient if alarms are not timely
transmitted to the nurse call system.
CAUTION
A nurse call system may only be connected by
trained electrical staff!
CAUTION
Check the nurse call system once a month.
Voltage
max. 40 V AC/DC
Switching capacity max. 500 mA
Contact
potential-free
025
1 2 3
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...