PC-AC
Pressure Control – Assist Control
Pressure-controlled, assisted ventilation
Every inspiratory effort of the patient on PEEP level
triggers a synchronized mandatory breath.
Set PC-AC with the buttons for the ventilation
parameters:
– Inspiration pressure »
P
insp
«
– Positive end expiratory pressure »
PEEP
«
– O
2
concentration »
FiO
2
«
– Sensitivity of »
Trigger
« (for synchronising
spontaneous breathing)
– Frequency »
f
«
– Inspiration time »
Ti
« or »
I:E
« (configurable, see
– Pressure rise time »
Ramp
«
z
Non-invasive ventilation with a hose system
with leak valve – using the »
Mask
« application
method, see page 75.
z
Setting alarm limits, see page 84.
037
Flow
Paw
Ti
PEEP
Pinsp
Ramp high
Ramp low
t
t
f
1
Trigger window
without spontane-
ous breathing
with spontaneous
breathing
524