
Instructions for Use Carina SW 3.n
97
Fault – Cause – Remedy
Battery indication not reliable
!
Battery indication is not relia-
ble.
Proceed the action on page 82.
If the problem persists call
DrägerService.
Blower overheated
(in log book »Blower over-
heated !!!«)
!!!
The blower has overheated.
The device is operating at
reduced output. Ventilation is
not optimal.
Make sure air is being supplied
to the device. If there is a free
supply of air: Use a replace-
ment device.
Call DrägerService.
Check settings
(in log book
»Check settings !!«)
!!
Loss of saved settings. Carina
operates with the default set-
tings.
Confirm with rotary knob. Use
a replacement device.
Call DrägerService.
Continuous Paw high
(in log book »Continuous Paw
high !!!«)
!!!
Expiratory valve obstructed.
Check expiratory valve.
Continuous Paw low
(in log book »Continuous Paw
low !!!«)
!!!
Note: This error is only sig-
nalled after 5 to 30 seconds.
Air supply blocked. The ventila-
tion function is not optimal.
Hose system blocked or leak-
ing.
Make sure there is a free sup-
ply of air. If there is a free sup-
ply of air: Check the hose
system, filters and other com-
ponents for free-flow and
leaks. Use a replacement
device. Call DrägerService.
Device failure
(in log book
»Device failure !!!«)
!!!
The device has a fatal techni-
cal problem. Ventilation has
failed.
Immediately disconnect the
patient from Carina and con-
tinue ventilation using a differ-
ent ventilator.
If the device cannot be turned
off, remove the O
2
hose from
the central supply and then
turn off the device.
Call DrägerService.
Disconnection/high leakage
!
Disconnection of the hose or
large leakage is detected.
After the leakage alarm, Carina
delivers a low constant flow at
PEEP level. When the leak has
been eliminated, Carina auto-
matically resumes operation at
the previous ventilation set-
tings.
Check hoses, filters and other
components to make sure the
connections are not leaking
and are secure.
Error message
Cause
Remedy
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...