Operation
82
Instructions for Use Carina SW 3.n
Improve displayed accuracy of the
charging capacity of the internal
battery
If the internal indication accuracy is not reliable
enough the following message will appear:
»
Battery indication not reliable !
«
Improving the display accuracy:
z
Connect power cable. The internal battery is
charging, the LED
flashes.
z
Charge internal battery completely, the LED
remains illuminated.
z
Disconnect the power cable from the unit. Con-
nect test lung and operate the Carina in manda-
tory ventilation mode (VC-SIMV, PC-AC or
PC-BIPAP) until the battery is completely dis-
charged. The LED
is no longer illuminated.
Carina stops ventilation.
Do not interrupt the discharging process!
z
Set the mains switch to off »
«.
z
Reconnect the power cable. The internal bat-
tery is charging, the LED
flashes. Charge
internal battery completely (at least 4 hours),
the LED
remains continually illuminated.
Do not interrupt the charging process!
If there is still no improvement, call DrägerService.
NOTE
If charging and discharging cycles are incom-
plete, the charging capacity of the internal battery
may be displayed inaccurately.
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...