Fault – Cause – Remedy
100
Instructions for Use Carina SW 3.n
MV high
!!
The minute volume has
exceeded the upper alarm
limit.
Check condition of patient,
check ventilation pattern, cor-
rect alarm limit if necessary.
Leakage in breathing system.
Ensure that breathing system
is leak-proof.
Device malfunction.
Disconnect the patient from
Carina and continue ventilation
immediately using a different,
separate ventilator.
Call DrägerService.
MV low
!!!
The lower alarm limit for minute
volume has been exceeded.
Check condition of patient,
check ventilation pattern, cor-
rect alarm limit if necessary.
Leakage in breathing system.
Ensure that breathing system
is leak-proof.
Device malfunction.
Disconnect the patient from
Carina and continue ventilation
using a different, separate ven-
tilator.
Call DrägerService.
O
2
calibration failed
!
O
2
calibration failed.
Call DrägerService.
Rescue Ventilation
(in log book »Rescue Ventila-
tion !!!«)
!!!
The device has a technical
problem and switches to emer-
gency operation. Ventilation is
not optimal.
Make sure there is a free sup-
ply of air to the inlet filter.
Remove any obstructions (for-
eign bodies, kinked hoses)
from the hose connections and
hose system.
Use a replacement device.
Call DrägerService.
Pilot line disconnected
!!!
A connection of the pilot hose
has become detached. (Only
for a system with an expiratory
valve) No ventilation is possi-
ble!
Make sure the pilot hose is
secure at both ends.
Check to avoid huge leakage.
P
max
limit reached
!
The VT measured is lower than
the VT set. The maximum per-
missible pressure range has
been reached.
Check the range of P
max
.
Check the setting for VT.
Set standby to shut-off
!!
Impermissible attempt to
switch off the device at the
main switch during ventilation.
Terminate ventilation first, then
switch off the device.
Error message
Cause
Remedy
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...