Instructions for Use Carina SW 3.n
71
Operation
Setting ranges
Ventilation parameter
Setting range
Pressure support
Δ
P
supp
[mbar]
LeakV: 2 to 40 (with PEEP)
ExpV: 2 to 50 (with PEEP)
PEEP [mbar]
3 to 20
Inspiratory O
2
concentration FiO
2
[%]
21 to 100
Trigger sensitivity Trigger
Normal/Sensitive
Maximum inspiration time Ti max [sec]
0.4 to 10.0
Ramp rise time Ramp [sec]
Auto/0.1 to 2.0
Apnoea ventilation Apn. Vent.
On/Off
Volume guarantee VG
On/Off
Pressure limitation P
max*
[mbar]
*
Only with apnoea ventilation or volume guarantee switched on
LeakV: 5 to 40
ExpV: 5 to 50