Operation
76
Instructions for Use Carina SW 3.n
In the »
Mask
« application method all the ventilation
modes are available for selection.
Selecting the »Mask« application method
z
Switch Carina to standby mode.
z
Press »
Select Menu
« button.
z
Press »
Config.
« button.
z
Press »
« button.
z
Press »
Method
« button.
z
Select and confirm »
Mask
« with rotary knob.
The status line shows »
Mask
«.
Setting ventilation parameters
z
Perform as described for the »
Tube
« applica-
tion method.
Automatic leak compensation of the triggers
Carina compensates leaks for the detection of a
trigger initiated by the patient. If leaks are higher
than 50 L/min, an alarm is given.
For further description see page 138.
Selecting the »Tube« application method
z
Switch Carina to standby mode.
z
Press »
Select Menu
« button.
z
Press »
Config.
« button.
z
Press »
« button.
z
Press »
Method
« button.
z
Select and confirm »
Tube
« with rotary knob.
The status line shows »
Tube
«.
z
Set alarm limits for the
»
Tube
«
application
method.
1
2
3
505
1
2
3
509
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...