Maintenance
110
Instructions for Use Carina SW 3.n
Maintenance of the internal battery
The battery has no memory effect.
If the battery is not used regularly it discharges.
Charge at regular intervals depending on ambient
temperature:
Every 3 months for –20 to +35 °C
Every month for +35 to +55 °C
Charge the battery more frequently at higher tem-
peratures.
Visual inspection of accessories
CAUTION
Even accessories designed to be reused (e.g.
after reprocessing) have a limited service life. Due
to a number of factors connected with handling
and reprocessing (e.g. disinfectant residues can
attack the material more intensely during auto-
claving), increased wear can occur and the serv-
ice life can be markedly shortened. These parts
must be replaced if signs of wear become visible,
such as cracks, deformation, discoloration, peel-
ing, etc.