
7ML19985KA02
SITRANS LG200 – INSTRUCTION MANUAL
Page 83
Te
chni
cal Ref
e
re
nce
By knowing the time between the first and second reflections, along with knowing the
upper layer dielectric constant, the thickness of the upper layer can be determined. In
order to properly process the reflected signals, the SITRANS LG200 is specified for those
applications where the thickness of the upper layer is greater than 2 inches. The
maximum upper layer is limited to the length of the Model 7ML1301-6 probe, which is
available in lengths from 60 to 610 cm (24 to 240").
Emulsion Layers
As emulsion (rag) layers can decrease the strength of the reflected signal, the SITRANS
LG200 should only be used in applications that have clean, distinct layers. Contact the
factory for application assistance.
Maximum Process Temperature Chart
Note:
The charts below are for guidance only:
Summary of Contents for 7ML1301-1
Page 1: ...Instruction Manual February 2009 LG200 sitrans ...
Page 10: ...Page 6 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SITRANS LG200 Notes ...
Page 59: ...7ML19985KA02 SITRANS LG200 INSTRUCTION MANUAL Page 55 SIMATIC PDM ...
Page 60: ...Page 56 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SIMATIC PDM Notes ...
Page 96: ...Page 92 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 Hazardous Installation ...