Page 18
SITRANS LG200 – INSTRUCTION MANUAL
7ML19985KA02
Specificat
ions
Approvals
North America
• General Purpose and Intrinsically Safe
CSA/FM Class I, Div. 1, Groups A, B, C, and D; Class II, Div.
1, Groups E, F, and G T4, Class III
• Non-Incendive
CSA Class I, Div. 2, Groups A, B, C, D, Class II, Div. 2,
Groups E, F, G; FM Class I, Div. 2, Groups A, B, C, D, Class
II, Div. 2, Groups F, G
• Explosion Proof
CSA/FM Class I, Div. 1, Groups B, C, and D; Class II, Div. 1,
Groups E, F, and G, T4; Class III
• Non-Incendive
CSA Class I, Div. 2, Groups A, B, C, D, Class II, Div. 2,
Groups E, F, G; FM Class I, Div. 2, Groups A, B, C, D, Class
II, Div. 2, Groups F, G
Europe
• General Purpose and Intrinsically Safe
ATEX II 1G EEx ia IIC T4
• Explosion Proof
ATEX II 1/2 GD EEx d [ia] IIC T6
• Non-sparking
ATEX II 3G EEx nA II/EEx nA (nL) IIC T4 to T6
Safety Integrity Level (SIL)
SIL-1
• Transmitter 7ML1300-xxAxx-xxxx Safe Failure Fraction (SFF) = 85.5%
SIL-2
• Transmitter 7ML1300-xxBxx-xxxx Safe Failure Fraction (SFF) = 91%
Note:
The device nameplate lists the approvals that apply to your device.
Further information can be found in SITRANS LG200 Functional Safety
Manual: 7ML19985LC01.
WARNING: Measured media inside vessel must be non-flammable
only. If media inside vessel is flammable, the explosion proof version is
required.
Summary of Contents for 7ML1301-1
Page 1: ...Instruction Manual February 2009 LG200 sitrans ...
Page 10: ...Page 6 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SITRANS LG200 Notes ...
Page 59: ...7ML19985KA02 SITRANS LG200 INSTRUCTION MANUAL Page 55 SIMATIC PDM ...
Page 60: ...Page 56 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 SIMATIC PDM Notes ...
Page 96: ...Page 92 SITRANS LG200 INSTRUCTION MANUAL 7ML19985KA02 Hazardous Installation ...