Page 54
SITRANS LG200 – INSTRUCTION MANUAL
7ML19985KA02
SI
MA
T
IC
PD
M
Loop Test
Allows you to input a simulated value in order to test the functioning of the mA
connections during commissioning or maintenance of the device.
To simulate a user-defined mA value:
1. Open the menu Device – Loop Test.
2. Select Other, enter the new value, and click on OK. The message ’Field Device
fixed at new value’ appears. Click on OK.
3. When you are ready to end simulation, select End and click on OK to return the
device to the original output level.
Self Test
Allows you to complete a self test function with the unit.
1. Open the menu Device –Self Test.
2. Select Yes, and the device will execute a self test. The LG200 will return to Run
mode upon test completion.