Standard/Left-Hand Type
Opposite/Right-Hand Type
(8) Base reference point after adjustment
[Advanced]
Set detailed sensor operation and print motion.
The setting items are as follows:
[Calibrate]
Adjust the media sensor level.
In instances of media detection malfunction, adjust the media sensor level (Gap and I-mark sensors).
The setting items are as follows:
[Auto-calibration]
Perform the auto-calibration for the selected media sensor.
[Auto-calibration] is not available if you have selected [Linerless] in the [Print Mode].
Gap + I-Mark
Adjusts both the Gap sensor and I-mark sensor.
Gap
Adjusts the Gap sensor.
I-Mark
Adjusts the I-mark sensor.
The procedure to auto-calibrate the media sensor is as follows:
1. Open the top cover.
2. Turn the head lock lever clockwise to unlock the print head, and then pull the feed lock latch to
unlock the feed roller and media sensor assembly.
3. Load the media.
Remove the label from the liner and set the liner so the I-mark is not above the I-mark sensor.
Summary of Contents for S84NX
Page 1: ...FX3 LX...
Page 124: ...Checking the Ink Side of the Ribbon 124 The ink is coated on the inner side Face in ribbon...
Page 320: ...Operations in Continuous Mode 320 Timing Chart With Head Check...
Page 363: ...363 8 Press the button again to change to Online mode You can print to check the print quality...
Page 371: ...371 6 Hold the adjust collar 2 in the set position and tighten the screw 3...
Page 457: ...457 9 Rotate the bearing clamp 3 onto the bearing and attach it with the screw 2...
Page 465: ...465 8 Push the center of the pressure plate 3 to latch it in place...
Page 469: ...469 9 Rotate the bearing clamp 4 onto the bearing and attach it with the screw 3...
Page 553: ...553 Printer Languages SBPL Includes XML support SZPL SDPL SIPL...
Page 588: ...Back Cover October 2022 Version 1...