BluePippin Operations Manual 460013 Rev F
12-4
The Elution Timer records the length of time of the elution. When elution is complete,
electrophoretic path will switch to the
either the Idle state (if the “End Run When Elution
is Completed” box is checked when the protocol was programmed. See
Figure 12.1
,
below) or ba
ck to the “Separate” separation channel (if the box was not checked). The
elution timer will show the value of the total elapsed elution time.
Run progress is monitored in the Protocol Status panel. The percent completion of the
run time is indicated by a progress bar.
The run time is determined by the “Run Time”
field in the protocol editor (see
Figure 12.1
, below). However, the run may end at a
sooner time if all elutions have been completed.
Important!
If the “End Run when Elution is Completed” is checked in the protocol editor, the run
may terminate before the progress bar reaches 100%
Figure 12.1
The Run Time field in the protocol editor. This
time value is used to determine the progress of the run.