![RS PRO RSDM-9060 User Manual Download Page 47](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059047.webp)
47
2W/4W Resistance Measurement
Measurement
type
2-wire OHM
Uses the standard Input HI-LO
terminals. Recommended for
measuring resistances larger than 1kΩ.
4-wire OHM
Compensates the test lead effect using
the 4W compensation terminals
(SENSE HI/LO terminals), in addition
to the standard Input HI-LO terminals.
Recommended for measuring sensitive
resistances smaller than 1kΩ.
Activate 2W or 4W
Measurement
Press the
Ω
2W key to activate 2W
resistance measurement.
Press the Shift
Ω
2W key to activate
4W resistance measurement.
+
2W/4W resistance
mode display
appears
The mode will switch to the selected resistance
mode immediately. Press the Shift
Ω
2W key on
the front panel as figure shown below.
2 or 4-Wire
OHM
Indicates 2W or 4W Resistance mode
5/s
Indicates the active refresh rate
Indicates Automatic range selection
Range: 100 Ω Indicates the available range of
Resistance
000.0651
Ω
Indicates the exact measured value
Summary of Contents for RSDM-9060
Page 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Page 2: ......
Page 5: ...5 Dimensions 409 Declaration of Conformity 410...
Page 141: ...141 EOM pulse width timing...
Page 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Page 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Page 409: ...409 Dimensions All dimensions are shown in millimeters...
Page 412: ...412 rspro com...