6.
Place the new fuse into the fuse holder. Gently
push the fuse downwards to make it firmly
fixed within the fuse holder.
7.
Reassemble the instrument properly followed
by connecting all the cables and cords.
8.
Fuse replacement is completed.
Battery Replacement
Beforehand
This chapter describes the procedure of battery
replacement in the front panel.
Before start, it is required to let a certified and
trained technician properly aware of potential
risks to disassemble instrument case. Unplug
power cord and disconnect external circuit from
the instrument before opening the case. Some of
the electrical connections are dynamic and even
available after powering off the instrument.
Consequently, Do disconnect all the inputs, cords
and cables before disassembling the instrument.
The steps to
replace battery
1.
Power off properly and disconnect all the test
leads, cables including power cord.
2.
Disassemble the instrument case in light of the
disassembling instructions.
3.
Find the battery (CR2032) on the main board,
which is perfectly located in the BT101 behind
the transformer.
Summary of Contents for RSDM-9060
Page 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Page 2: ......
Page 5: ...5 Dimensions 409 Declaration of Conformity 410...
Page 141: ...141 EOM pulse width timing...
Page 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Page 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Page 409: ...409 Dimensions All dimensions are shown in millimeters...
Page 412: ...412 rspro com...