![RS PRO RSDM-9060 User Manual Download Page 303](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059303.webp)
303
CALCulate Commands
CALCulate:CLEar[:IMMediate]
Clears all of the compare results, statistic calculation value,
histogram calculation value, and measurement value.
Parameter: <None>
Example: CALC:CLE:IMM
CALCulate:DATA?
Returns uncalculated original measurement.
CALCulate:FUNCtion
Sets the Advanced function.
Parameter: OFF | HOLD | DB | DBM | LIM | MXB | INV | REF
Example: CALC:FUNC DB
Sets the Advanced function to DB mode
CALCulate:FUNCtion?
Returns the current Advanced function.
Return parameter: OFF | HOLD | DB | DBM | LIM | MXB | INV |
REF
CALCulate:HOLD:REFerence
Sets the percentage threshold for the Hold function.
Parameter: <NRf> (0.01, 0.1, 1, 10)
Example: CALC:HOLD:REF 10
Sets the hold percentage to 10%.
CALCulate:HOLD:REFerence?
Returns the percentage threshold from the Hold function.
Return parameter: 0.01 | 0.1 | 1 | 10
CALCulate:STATe
Turns the Advanced function on/off.
Parameter: 0 | 1 | ON | OFF
Example: CALC:STAT OFF
Turns the Advanced function off.
CALCulate:STATe?
Returns the status of the Advanced function.
Return Parameter: 0 | 1, 1=ON, 0=OFF
CALCulate:AVERage:ALL?
Returns all of the statistic calculation values.
Return parameter: average, standard deviation, minimum,
maximum values.
Summary of Contents for RSDM-9060
Page 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Page 2: ......
Page 5: ...5 Dimensions 409 Declaration of Conformity 410...
Page 141: ...141 EOM pulse width timing...
Page 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Page 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Page 409: ...409 Dimensions All dimensions are shown in millimeters...