Measurement Method: Selectable 4-wire or 2-wire ohms. Current source
referenced to Input LO Terminal
DC Current
Range
Shunt
Burden Voltage
100
μ
A
100
Ω
<0.011 V
1 mA
100
Ω
<0.11 V
10 mA
1
Ω
<0.04 V
100 mA
1
Ω
<0.4 V
1 A
0.1
Ω
<0.7 V
3 A
0.1
Ω
<2 V
10 A
10m
Ω
<0.5 V
Input Protection
External 3.15 A, 500 V fuse for 3 A
Internal 6 A, 1 kV fuse for 3 A
Internal 12 A, 1 kV fuse for 10 A
Reading Rate
(Readings/sec)
Speed
Digits
DCV
DCI
2W/4W-Resistance
5 /s
,
20 /s
,
60 /s
,
100 /s
6 ½
400 /s
,
1.2 k /s
,
2.4 k /s
5 ½
4.8 k /s
,
7.5 k /s
,
10 k/s
4 ¼
Speed
Digits
Continuity
Diode
60 /s
6 ½
100 /s
5 ½
400 /s
4 ¼
[1]
DC Specification: In addition to the availability that requires warm-up of 60
minutes, it must be set in 5/s speed rate (60/s speed rate for Continuity and
Diode), A-Zero on.
[2]
The entire range of measurement will pass the set range by 20% except the
tests of 1000 V DC, 3 A DC, 10 A DC and diode.
[3]
This specification applies to 4-wire resistance measurement, whilst it requires
using "REL" function for offset on 2-wire resistance measurement. 2-wire
resistance measurement will cause additional error of 0.2
Ω
if REL function is
not executed.
[4]
This specification applies to the voltage measured from input terminal. 1 mA
test current is the typical value. The change of current source leads to the
variation in buck of diode junction.
[5]
Accuracy is ± (DC Input ac DC Reference accuracy), where Input
accuracy = DC Voltage accuracy for the Input HI to LO (in % of the Input
voltage), and Reference accuracy = DC Voltage accuracy for the HI to LO
(Sense) Reference (in % of the Reference voltage).
[6]
The 10 A range of measurement is available for the terminals on front panel
only. Due to power factor resulting in temperature rise, 2 mA increment per
one ampere when input is greater than 5 A.
Summary of Contents for RSDM-9060
Page 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Page 2: ......
Page 5: ...5 Dimensions 409 Declaration of Conformity 410...
Page 141: ...141 EOM pulse width timing...
Page 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Page 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Page 409: ...409 Dimensions All dimensions are shown in millimeters...
Page 412: ...412 rspro com...