![RS PRO RSDM-9060 User Manual Download Page 400](http://html1.mh-extra.com/html/rs-pro/rsdm-9060/rsdm-9060_user-manual_1496059400.webp)
400
rspro.com
1.000000 A < 0.7 V 0.050 + 0.006 0.080 + 0.010 0.100 + 0.010 0.0050 +0.0010
3.000000 A < 2.0 V 0.180 + 0.020 0.200 + 0.020 0.200 + 0.020 0.0050 +0.0020
Accuracy Specifications: ± ( % of r % of range )
Continuity
Range
[2]
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
1 k
Ω
0.003 + 0.030 0.011 + 0.030 0.014 + 0.030 0.001 + 0.002
Accuracy Specifications: ± ( % of r % of range )
Diode Test
[4]
Range
[2]
24 Hour
TCAL ± 1 ˚C
90 Day
TCAL ± 5 ˚C
1 Year
TCAL ± 5 ˚C
Temperature
Coefficient/˚C
5 V
0.003 + 0.030 0.011 + 0.030 0.014 + 0.030 0.0010 + 0.0020
Accuracy Specifications: ± ( % of r % of range )
DCV Ratio
[5]
Accuracy Specification: ± (DC Input ac DC Reference accuracy)
Measuring Characteristics
DC Voltage
Input Resistance Range
100 mV
10 M
Ω
or >10 G
Ω
Selectable
1 V
10 V
100 V
10 M
Ω
±1%
1000 V
Input Bias
30 pA (Typ, 25°C)
Input Protection 1000 V on all ranges
Measurement Method: Sigma-delta A/D Converter
Resistance
Max. Lead
Resistance
10% of range per lead for 100
Ω
, 1 k
Ω
ranges. 1 k
Ω
per lead on all other ranges.
Input Protection 1000 V on all ranges
Measurement Method: Selectable 4-wire or 2-wire ohms. Current source
Summary of Contents for RSDM-9060
Page 1: ...User Manual RSDM 906X series Stock number 1919720 RSDM 9061 1919721 RSDM 9060 EN rspro com...
Page 2: ......
Page 5: ...5 Dimensions 409 Declaration of Conformity 410...
Page 141: ...141 EOM pulse width timing...
Page 160: ...160 rspro com SYSTEM FIRMWARE View System Info 161 Firmware Update 162...
Page 211: ...211 SCREENHOT LOG Capture 212 Save Reading 215...
Page 409: ...409 Dimensions All dimensions are shown in millimeters...
Page 412: ...412 rspro com...