Rosemount 5300 Series with HART® to Modbus® Converter
16.
When upload is finished, select
Diagnostics
in the
Tools
menu.
17.
Click
Device Errors
and check for “Checksum”.
18.
If it is on the list, choose the
Factory Settings
option from the
Tools
menu.
19.
Select
All
and click
OK
.
20.
Select
Yes
.
Note
An error message might be displayed when performing the Reset to Factory Settings
operation. The operation has been successful if the checksum error has been cleared.
21.
Select
Restart
in the
Tools
menu to restart the HMC.
22.
The checksum error should no longer be present (select
Diagnostics
in the
Tools
menu to verify, see Step
). If it still persists, follow the next steps.
Checksum