112
Reference Manual
00809-0100-4530, Rev DD
May 2016
Analog output is not available for F
OUNDATION
Fieldbus.
Typically, the
Primary Variable
(PV) is configured to be product level, interface
level, or volume. Other variables like product distance, interface distance, upper
product thickness, etc. are available as well.
Specify the analog output range by setting the
Lower Range Value
(4 mA) and the
Upper Range Value
(20 mA) to the desired values.
The alarm mode specifies the output state when a measurement error occurs.
“Analog output (HART)” on page 94
for information on analog output
configuration.
Finish configuration wizard
11.
Finish Configuration Wizard.
This is the last window in the Configuration Wizard concluding the basic
configuration. The current configuration can be changed at any time by using the
setup windows (General, Tank, Output etc., see
“Using the setup functions” on
The setup windows contain further options not available in the Configuration
Wizard.
Select
Finish
and continue with the next step in the guided setup.