7.22.7
Analog output status
Analog output status messages that may appear on the integral display, on the Field
Communicator, or in RRM are shown in
.
Note
Analog output status messages may not appear on the integral display while the
Rosemount 5300 Series Transmitter runs in HART 7 multidrop mode.
Table 7-10. Analog Output Status
Message
Description
Action
Not connected
Analog output hardware is not connected.
Alarm Mode
The analog output is in Alarm Mode.
Check Error and Warning Messages to
find the reason for the Alarm.
Saturated
The analog output signal value is saturated,
i.e. equal to the saturation value.
No action needed.
Multidrop
The transmitter is in Multidrop Mode. The
analog output is fixed at 4 mA.
This is the normal setting when a device is
used in Multidrop configuration.
Fixed Current mode
The analog output is in fixed current mode.
This mode is used when calibrating the
Analog Output channel.
PV out of Limits
The Primary variable is out of range.
Check the Upper and Lower Range Value
Span Too Small
The configured span is too small.
Check the Upper and Lower Range Value.
Invalid Limits
The given Upper and Lower Sensor Limits are
invalid.
Check that the difference between the
Upper and Lower Sensor Limits is greater
than the Minimum Span.