57
Reference Manual
00809-0100-4530, Rev DD
May 2016
Coaxial probe
The coaxial probe can be guided by a
tube welded on the tank bottom.
Tubes are customer supplied. Make
sure that the probe can move freely in
order to handle thermal expansion.
The measurement accuracy will be
reduced close to the tube opening.
Rigid twin lead probe
The rigid twin lead probe can be
secured to the tank wall by cutting the
center rod and putting a fixture at the
end of the outer rod.
The fixture is customer supplied. Make
sure the probe is only guided and not
fastened in the fixture to be able to
move freely for thermal expansion.
Flexible single lead probe
The flexible single lead probe itself can
be used for anchoring. Pull the probe
rope through a suitable anchoring
point, (e.g. a welded eye) and fasten it
with a chuck.
The length of the loop will add to the
Blind Zone.The location of the clamps
will determine the beginning of the
Blind Zone. See section
for
further information on Blind Zones.
The probe length should be
configured as the length from the
underside of the flange to the top
clamp.
Drain
Ø 0.3 in.
(8 mm)