7-6
IMSBR10-E1
7.4 Setting the RJC Function on Channels Set
to TC Input
This section explains the details of setting RJC on channels set to TC input.
Basic Setting mode cannot be entered when recording is in progress or when computation is in progress on
models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic Setting mode.
3.
Press the
key or
SHIFT
+
key to select
RJC
and then press the
key.
4.
Set each item and press the
key.
Use the
key or
SHIFT
+
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the
ESC
key in the middle of the operation, the settings entered up to then are cancelled,
and the display returns to a higher level menu.
5.
When the
Setting complete
screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the
ESC
key.
Set the channel range.
Select the RJC mode of the
thermocouple input.
Set the compensation voltage when
using the external RJC function.
The new setting takes effect.
ESC/?
First channel
Last channel
(When RJC is set to External)
01-01 Channel
Setting complete
Basic=RJC
CH=01-01
RJC=Internal
Volt= 0
Applying the Changes and Returning to Operation Mode
Press the
key or
SHIFT
+
key to select
End
and then press the
key. Press the
key to select
Store
and then press the
key. The changes are applied, and the screen returns to
Operation mode. If you select
Abort
and press the
key, the changes are discarded, and the screen
returns to Operation mode. Press the
ESC
key to return to the
Basic=
screen.