11-9
IMSBR10-E1
Maintenance
11
11.7 Recommended Replacement Periods for
Worn Parts
To preserve the reliability of the recorder and to use the recorder in good condition for an extended time, it is
recommended that periodic replacements be made on parts.
The table below shows the recommended replacement period for expendable parts.
The replacement period shown here applies when the recorder is used under standard operating conditions. For
the actual replacement period, consider the actual conditions of use.
Replacement of parts other than the chart paper, pen, ribbon cassette, and internal light LED will be contact your
nearest RKC INSTRUMENT dealer when such replacement is necessary.
Note
The replacement period of the display and the internal light LED is the half life of the brightness. The
deterioration of brightness varies depending on the condition of use, and its determination is subjective.
Consider these facts for determining the actual replacement period.
Pen Model
Note
41 days
2 km
B8802CA
MOTOR ASSY
5 years
Item
Replacement
Period
Part Name
Part No.
Z-fold chart paper
CHART
R-100EX
When used at 20 mm/h
33 days
When used at 25 mm/h
Pen servo
Lever
Plotter motor
Plotter carriage
Chart paper feed
motor
Display
Plotter pen
Felt pen
PEN ASSY
B9902AM
B9902AN
B9902AP
B9902AQ
Red
Green
Blue
Violet
At a pen speed of
10 cm/s
100,000
characters
PEN ASSY
B9902AR
When printing continuously
5 years*
DISPLAY ASSY
5 years
MOTOR ASSY
B9905GS
5 years
CARRIAGE ASSY B9905NB
For the X-axis
5 years
B9900RG
Lever ASSY
B9900RH
For the plotter
5 years
SERVO ASSY
B8802EF
Shared by all pens
(excludes the pen arm ASSY)
Quantity
Used
1
1 each
1
1
1
1
Internal light LED
2 years*
LED ASSY
B8800CR
2
1
1
1 to 4
* The half life of the brightness at the factory default brightness setting.