4-14
IMSBR10-E1
4.4 Function Setup Guide
This section explains the settings necessary to use various functions of the recorder. Read the section
corresponding to the function you wish to use.
Note
This section contains all the settings related to each item. If the desired setting is the same as the default
value, you do not have to set it.
Item
Description
Reference
Section
Date/Time setting
Use Clock in Setting mode
5.5
DST
Sets the date/time for switching between DST and standard time using
6.13
Aux > DST in Setting mode.
Setting initialization
Use Initialize in Basic Setting mode to initialize the settings of Setting mode
7.17
and Basic Setting mode to their default values.
Measuring input functions
Item
Description
Reference
Section
Input range of each channel
Set Range in Setting mode.
5.1
1-5V
• Unit
5.3
Use Unit in setting mode to set the unit after scaling.
• Low-cut
7.15
Use Personalize > Add function > 1-5V low-cut in Basic Setting mode and
select Use or Not.
5.1
If Use is selected, turn On/Off the low-cut function using Range in Setting mode.
If Not is selected, the Low-cut item does not appear in the Range setting.
Scaling
• Unit
5.3
Use Unit in setting mode to set the unit after scaling.
Square Root Computation
• Unit
5.3
Use Unit in setting mode to set the unit after scaling.
• Low-cut
7.15
Use Personalize > Add function > SQRT low-cut in Basic Setting mode
and select Use or Not.
If Use is selected, set the low-cut value using Range in Setting mode.
5.1
If Not is selected, the Low-cut item does not appear in the Range setting.
Bias
Use Personalize > Add function > Bias in Basic Setting mode and select Use
7.15
or Not.
6.12
If Use is selected, set the bias value that is added to the input using Bias in
Setting mode. If Not is selected, the Bias item does not appear.
Filter (pen model)
Use Filter in Basic Setting mode and select Use or Not.
7.12
If Use is selected, set the filter time constant using Aux > Filter in Setting mode.
6.2
If Not is selected, the Aux > Filter item does not appear.
Moving average (dot model)
Use Moving_AVE in Basic Setting mode and select Use or Not.
7.11
If Use is selected, set the number of samples of moving average using Aux >
6.3
Moving_AVE in Setting mode.
If Not is selected, the Aux > Moving_AVE item does not appear.
RJC of TC input
Use RJC in Basic Setting mode to select whether to use the recorder’s
7.4
RJC function or an external RJC function.
Burnout detection function of TC input and 1-5V input
7.3
Use Burnout in Basic Setting mode to set the burnout detection function for
each channel.
Integration time of the A/D converter
7.2
Use A/D in Basic Setting mode to set the integration time of the A/D converter.
Temperature Unit
Select the temperature unit using Temperature in Basic Setting mode.
7.20
Summary of Contents for SBR-EW180
Page 1: ...IMSBR10 E1 SBR EW180 Recorder User s Manual...
Page 2: ......
Page 14: ......
Page 233: ...3 2002...